General specifications, 1 standard specifications, General specifications -1 – Yokogawa EJA220A User Manual

Page 58: 1 standard specifications -1

Advertising
background image

IM 01C21C01-01E

10-1

10. GENERAL SPECIFICATIONS

10. GENERAL SPECIFICATIONS

10.1

Standard Specifications

Refer to IM 01C22T02-01E for F

OUNDATION

Fieldbus communication type and IM 01C22T03-
00E for PROFIBUS PA communication type
marked with “

᭛”.

Performance Specifications

See General Specifications sheet, GS 01C21C01-
00E.

Functional Specifications

Span & Range Limits:

M

H

Measurement

Span

and Range

Span

Range

Span

Range

inH

2

O

(/D1)

4 to 400

–400 to 400

20 to 2000

–2000

to 2000

mbar

(/D3)

10 to 1000

mmH

2

O

(/D4)

100 to 10000

–1000

to 1000

50 to 5000

–5000

to 5000

–10000

to 10000

0.05 to

5 kgf/cm

2

–5 to

5 kgf/cm

2

kPa

1 to 100

–100

to 100

5 to 500

–500

to 500

T1001.EPS

URL is define as the Upper Range Limit from the table
above.

Zero Adjustment Limits:

Zero can be fully elevated or suppressed, within
the Lower and Upper Range Limits of the capsule.

External Zero Adjustment “

”:

External zero is continuously adjustable with
0.01% incremental resolution of span. Span may
be adjusted locally using the digital indicator with
range switch.

Output “

”:

Two wire 4 to 20 mA DC output with digital
communications, linear or square root program-
mable. BRAIN or HART FSK protocol are superim-
posed on the 4 to 20 mA signal.

Failure Alarm:

Output status at CPU failure and hardware error;
Up-scale: 110%, 21.6 mA DC or more(standard)
Down-scale: -5%, 3.2 mA DC or less

-2.5%, 3.6 mA DC or less(Optional

code /F1)

Note: Applicable for Output signal code D and E

Damping Time Constant (1st order):

The sum of the amplifier and capsule damping
time constant must be used for the overall time
constant. Amp damping time constant is adjustable
from 0.2 to 64 seconds.

Time Constant (approx. sec)

0.4 0.4 0.4 0.4

M H

M

H

Capsule (Silicone Oil)

EJA210A

EJA220A

Ambient Temperature Limits:

* Safety approval codes may affect limits.

–40 to 85°C (–40 to 185°F)
–30 to 80°C (–22 to 176°F) with LCD Display

Process Temperature Limits:

* Safety approval codes may affect limits.

–40 to 120°C (–40 to 248°F)

Working Pressure Limits (Silicone Oil)

2.7 kPa abs (0.38 psia) to flange rating
(see Figure 1.)

Installation

Supply & Load Requirements “

”:

* Safety approvals can affect electrical requirements.

See Section 6.6, ‘Power Supply Voltage and Load
Resistance.’

Supply Voltage “

”:

10.5 to 42 V DC for general use and flameproof type
10.5 to 32 V DC for lightning protector (Optional code/A)
10.5 to 30 V DC for intrinsically safe, Type n,
nonincendive, or non-sparking type
Minimum voltage limited at 16.4 V DC for digital
communications, BRAIN and HART

EMC Conformity Standards“

”:

,

EN61326-1 Class A, Table 2 (For use in industrial

lications)

EN61326-2-3

Communication Requirements “

”:

BRAIN
Communication Distance;

Up to 2 km (1.25 miles) when using CEV polyethyl-
ene-insulated PVC-sheathed cables.
Communication distance varies depending on type
of cable used.

Load Capacitance;

0.22 µF or less (see note)

Load Inductance;

3.3 mH or less (see note)

Input Impedance of communicating device;

10 kΩ or more at 2.4 kHz.

Note: For general-use and Flameproof type.

For Intrinsically safe type, please refer to
‘Optional Specifications.’

Advertising