13) zero point adjustment, 13) zero point adjustment -13 – Yokogawa EJA220A User Manual

Page 45

Advertising
background image

IM 01C21C01-01E

8-13

8. BRAIN TERMINAL BT200 OPERATION

• Example 2: When the higher range value is to

be changed to 10 kPa with the

present setting of 0 to 30 kPa, take
the following action with an input
pressure of 10 kPa applied.

FEED

NO

OK

SET
H11:AUTO HRV
10.000 kPa

DATA

DIAG

PRNT

ESC

PARAM
H10:AUTO LRV
0 kPa
H11:AUTO HRV
10.000 kPa
H60:SELF CHECK
GOOD

ESC

SET
H11:AUTO HRV
30 kPa
+ 30

The lower range value is not

changed, so the span changes.

Parameter C22 is changed at the

same time.

Press the key twice.

The higher range value is changed

to 10 kPa.

Press the (OK) key.

F0828.EPS

(13)

Zero Point Adjustment
(J10: ZERO ADJ, J11: ZERO DEV,
J20: EXT ZERO ADJ)

The DPharp supports several adjustment methods.
Select the method best suited for the conditions of your
application.
Note that output signal can be checked by displaying
parameter A10:OUTPUT (%) on the BT200.

Adjustment Method

Description

Using the BT200

Set the present input to 0%.

Using the external
zero-adjustment
screw

Adjust zero point using the zero-
adjustment screw on the transmitter.
This permits zero adjustment without
using the BT200. Accurately adjust
the output current to 4 mA DC or
other target output value using an
ammeter that accuratly reads output
currents.

Adjust output to the reference value
obtained using other means.

Adjust for 0% output at input level of
0%.

If the input level cannot easily be
made 0% (because of tank level,
etc.), adjust output to the reference
value obtained using other means,
such as a sight glass.

T0809.EPS

(a) Follow the procedure below when setting the

present output to 0% (4 mA).

FEED

NO

OK

SET
J10:ZERO ADJ
0.0 %

CLR

ESC

SET
J10:ZERO ADJ
0.0 %
+ 000.0

Press the key twice.

Zero adjustment is completed.

Press the (OK) key.

F0829.EPS

Output is 0.5%.

A10:OUTPUT (%)
0.5 %

Output is 0%.

A10:OUTPUT (%)
0.0 %

(b) In tank level measurement, if the actual level

cannot be brought to zero for zero adjustment, then
the output can be adjusted to correspond to the
actual level obtained using another measuring
instrument such as a glass gauge.

Present level: 45%
Present output: 41%

F0830.EPS

100%

0%

Present level

45%

(b)-1 Follow the procedure below to use J10: ZERO

ADJ.

CLR

ESC

SET
J10:ZERO ADJ
0.0 %
+ 045.0

Enter the present actual level, 45%.

Press the key twice.

The output is changed to 45%.

F0831.EPS

Present output is 41.0%.

A10:OUTPUT (%)
41.0 %

A10:OUTPUT (%)
45.0 %

Advertising
This manual is related to the following products: