Yokogawa EJX440A User Manual

Page 23

Advertising
background image

IM 01C25T03-01E

3-15

3. OPERATION

When using BT200, the output signal can be adjusted
either in % or pressure unit. The unit can be selected
by the parameter J09: ADJ UNIT. Output signal can be
changed by displaying parameter A10: OUTPUT for %
or J10: OUTPUT for pressure unit.

This section discribes the zero adjustment procedure by
using the pressure unit.

a-1. Zeroing

Setting the parameter of J11: P ZERO ADJ carries out
the zero adjustment and automatically sets the applied
“0” input values to the transmitter’s output value of
zero, keeping the span constant. Use this setting when
the LRV is known to be 0 kPa.

FEED

NO

OK

SET
J11:P ZERO ADJ
0.00000 kPa

CLR

DEL

ESC

SET
J11:P ZERO ADJ
0.00000 kPa
+ 0

A pressure of 0 kPa is applied.

Press the key twice

after the pressure has become

stable.

Zero adjustment is completed.

Press the (OK) key.

F0331.EPS

Transmitter measures pressure of

0.03585 kPa.

Transmitter measures pressure of

0.00000 kPa.

A11:PRES
0.03585 kPa

A11:PRES
0.00000 kPa

a-2. Level Adjustment

The zero adjustment by the parameter of J11: P ZERO
ADJ calibrates the transmitter output corresponding to
the actual tank level. To perform this adjustment, first
use a glass gauge or the like to determine the actual
tank level, then enter the correct data as shown below.

F0332.EPS

25.00 kPa

0.00 kPa

Actual level
13.50 kPa

DPharp span:

0 to 25.00 kPa

Actual level:

13.50 kPa

Transmitter output: 13.83 kPa

DPharp

CLR

DEL

ESC

SET
J11:P ZERO ADJ
0.00000 kPa
+ 045.0
+ 13.5

Enter the present actual level,

13.5 kPa.

Press the key twice.

The measured pressure is changed

to 13.5 kPa.

F0333.EPS

Transmitter measures present

pressure of 13.83 kPa.

A11:PRES
13.83 kPa

A11:PRES
13.5000 kPa

a-3. Using External Zero-adjustment Screw

This method permits zero adjustment without the
BT200. Use a slotted screwdriver to turn the zero-
adjustment screw. See the hardware manual for details.

Note that the parameter of J55: EXT ZERO ADJ must
be

ENABLE

to perform this adjustment.

Follow the procedure below to enable or inhibit zero
point adjustment from the zero-adjustment screw on
the transmitter.
This is set to

ENABLE

when the instrument is

shipped.

• Example: Inhibiting zero adjustment by the

external zero-adjustment screw.

ESC

SET
J55:EXT ZERO ADJ
ENABLE
< ENABLE >
< INHIBIT>

Use the or key to

select INHIBIT.

Press the key twice to

enter the setting.

F0335.EPS

Then press the (OK) key.

b. Full Sensor Trim

(J11: P ZERO ADJ, J12: P SPAN ADJ,
J15: P ZERO DEV, J16: P SPAN DEV)

Full sensor trim is carried out with a series of the
procedure of J11: P ZERO ADJ and J12: P SPAN
ADJ. Also, you can manually perform the trimming
procedure by using J15: P ZERO DEV and J16: P
SPAN DEV.

The full sensor trim is a two-point adjustment, and the
lower point adjustment should always be performed
before the upper point adjustment in order to maintain
the pitch between the zero and 100% points within the
calibration range.

In the manual method, the reference pressure should
also be applied to the transmitter at both lower and
upper point of trim ends. Without the reference
pressure, J15: P ZERO DEV and J16: P SPAN DEV
may not represent the correct value of adjustment point
for each.