Yokogawa EJX440A User Manual

Page 35

Advertising
background image

IM 01C25T03-01E

5-3

5. PARAMETER SUMMARY

No.

Parameter name

Item

R/W

Content

Applicable model

Upload

data

F

P

L

Default value

J ADJUST

Adjusting

data

J09

ADJ UNIT

Pressure adjusting unit select

W

% or PRES UNIT

J10

ADJ PRES

Adjustment reference pressure

R

Unit specified in J09

J11

P ZERO ADJ

Automatic zero adjustment

W

Ϫ

32000 to 32000, unit specified in J09

J12

P SPAN ADJ

Automatic span adjustment

W

Ϫ

32000 to 32000, unit specified in J09

J15

P ZERO DEV

Manual zero adjustment

W

Ϫ

32000 to 32000, unit specified in J09

J16

P SPAN DEV

Manual span adjustment

W

Ϫ

32000 to 32000, unit specified in J09

J20

ADJ SP

Adjustment reference pressure of SP

R

Unit specified in J09

J21

SP ZERO ADJ

Automatic SP zero adjustment

W

Ϫ

32000 to 32000, unit specified in J09

J22

SP SPAN ADJ

Automatic SP span adjustment

*

2

W

Ϫ

32000 to 32000, unit specified in J09

J25

SP ZERO DEV

Manual SP zero adjustment

W

Ϫ

32000 to 32000, unit specified in J09

J26

SP SPAN DEV

Manual SP span adjustment

W

Ϫ

32000 to 32000, unit specified in J09

J40

OUTPUT 4mA

4 mA adjustment

W

Ϫ

10.000 to 10.000%

J41

OUTPUT 20mA

20 mA adjustment

W

Ϫ

10.000 to 10.000%

J45

AMP TEMP

Amplifier temperature

R

Unit specified D40

J50

ADJ WHO

Adjustment information

W

8 alphanumeric characters

J51

ADJ DATE

Adjustment information

W

16 alphanumeric characters

J52

ADJ LOC

Adjustment information

W

8 alphanumeric characters

J53

ADJ DESC

Adjustment information

W

16 alphanumeric characters

J55

EXT ZERO ADJ

External zeroing permission

W

INHIBIT or ENABLE

J56

CLEAR ADJ

Clear adjustment

W

---, PRES, SP, 4-20mA, or ALL

J60

SELF CHECK

Self-diagnostics

R

See A60

K TEST

Test

parameters

K10

OUTPUT X %

Test output % setting

W

Within a range between D20 and D21

K40

DO TEST

Test contact output

W

OFF or ON

K50

TEST KEY1

Special maintenance parameter

W

K51

TEST KEY2

Special maintenance parameter

W

K52

TEST KEY3

Special maintenance parameter

W

K53

TEST KEY4

Special maintenance parameter

W

K60

SELF CHECK

Self-diagnostics

R

See A60

M

DEVICE INFO

Device information

M10 SERIAL NO.

Serial number

R

M11 MFTR. DATE

Manufactured date

R

M12 EXTRA NO.

Customization number

R

M15 SOFT REV

Software revision

R

M16 BRAIN REV

BRAIN protocol revision

R

M17 MEMO1

Memo

W 16

alphanumeric

characters

M18 MEMO2

Memo

W 16

alphanumeric

characters

M19 MEMO3

Memo

W 16

alphanumeric

characters

M20 ISOL MATL

Capsule material

W

M21 FILL FLUID

Fill fluid

W

M22 GASKET MATL

Gasket material

W

M23 PRO CON MATL Flange material

W

M24 D-VENT MATL

Vent plug material

W

M25 PRO CON TYPE Process connection type

W

M26 RS ISOL MATL

Remote seal material

W

M27 PRO CON SIZE

Flange size

W

M28 NUM RS

Number of remote seal

W

M29 RS FILL FLUID

Fill fluid of remote seal

W

M30 RS TYPE

Remote seal type

W

M50 MS CODE 1

Model and suffix code 1

W

M51 MS CODE 2

Model and suffix code 2

W

M52 MS CODE 3

Model and suffix code 3

W

M53 MS CODE 4

Model and suffix code 4

W

M54 MS CODE 5

Model and suffix code 5

W

M55 MS CODE 6

Model and suffix code 6

W

M60 SELF CHECK

Self-diagnostics

R

See A60

PRES UNIT

0.00000 kPa

100.000 kPa

100.000 kPa

100.000 kPa

0.00000 MPa

0.00000 MPa

0.00000 MPa

0.000%

0.000%

0 degC

ENABLE

---

0.00%

OFF

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

*

1: R/W: R = Read only, W = Read & Write

*

2: The default value shows MWP (Maximum working pressure) of the capsule.

Since the working pressure limit varies according to the Model, refer to the General Specifications section in each user’s manual.

*

1

Advertising