3 displaying data using the bt200, 1 displaying measured data, 2 display transmitter model and specifications – Yokogawa EJX440A User Manual

Page 29: Displaying data using the bt200 -21, Displaying measured data -21, Display transmitter model and specifications -21

Advertising
background image

IM 01C25T03-01E

3-21

3. OPERATION

(21)

Adjustment Information and User Memo
Fields (J50: ADJ WHO, J51: ADJ DATE,
J52: ADJ LOC, J53: ADJ DESC, M17 to
M19: MEMO1 to MEMO3)

This feature provides four fields for instrument
adjustment information at maintenance: inspection
date, inspector, location, and description. Also three
user memo fields are provided, each holding up to 16
alphanumeric characters.

• Example: Save an inspection date of October

21, 2003.

DATA

DIAG

PRNT

ESC

SET
J51:ADJ DATE
MM-DD-YYYY
10-21-2003

ESC

CODE

CAPS

CLR

PARAM
J50:ADJ WHO

J51:ADJ DATE

J52:ADJ LOC

Set “10-21-2003” in the order of

month, day, and year.

Press the key twice to

enter the setting.

F0337.EPS

3.3 Displaying Data Using the

BT200

3.3.1 Displaying Measured Data

The BT200 can be used to display measured data.

The measured data is updated automatically every
seven seconds. In addition, the display can be updated
to the present data value at any time by pressing the

(DATA)

key. For parameters associated with the

display of measured data, see chapter 5 Parameter
Summary.

• Example: Display output.

DATA

DIAG

PRNT

ESC

HOME

SET

ADJ

ESC

F0338.EPS

Data is updated automatically
at 7-second intervals.

Display “A10: OUTPUT.”

communi

PARAM
A10:OUTPUT
0.0 %
A11:PRES
0.0000 kPa
A15:OUTPUT mA
4.000 mA

PARAM
A10:OUTPUT
0.0 %
A11:PRES
0.0000 kPa
A15:OUTPUT mA
4.000 mA

MENU
A:DISPLAY
B:SENSOR TYPE
C:BASIC SETUP
D:AUX SET1
E:AUX SET2
G:ALARM SET

3.3.2

Display Transmitter Model and

Specifications

The BT200 can be used to display the model and
specifications of the transmitter.

• Example: View transmitter model name.

HOME

SET

ADJ

ESC

DATA

DIAG

PRNT

ESC

PARAM
B10:MODEL
EJX110 M
B11:STYLE NO.
1.00
B20:PRES LRL
– 98.07 kPa

Press .

F0339.EPS

For the associated
parameters, see Chapter 5,
Parameter Summary.

MENU
A:DISPLAY
B:SENSOR TYPE
C:BASIC SETUP
D:AUX SET1
E:AUX SET2
G:ALARM SET