Calibration, Calibration -1 – Yokogawa PH72 Personal pH/ORP Meter User Manual

Page 28

Advertising
background image

IM 12B03D02-01E

4-1

4. Calibration

4. Calibration

Calibration using standard solutions means to measure the pH value of a certified
standard solution and to adjust the pH meter so it reads the same value as the certified
value of the standard solution. The PH72 meter can be calibrated automatically or
manually.

The PH72 meter must be calibrated before measurement if:

• it has been stored for a long period;

• the electrode has been cleaned; or

• otherwise necessary.

Precautions

(1) Use certified standard solutions. Using deteriorated standard solutions will result

in inaccurate calibration. Standard solutions are available as spare parts (see
Section 1.8).

Pour a portion of the standard solution into a calibration bottle (supplied) and use
for calibration. Do not reuse the portion. Discard it and use a fresh portion for
another calibration.

(2) Do not press

CAL

key unless calibration is needed. Otherwise, saved calibration

results may be changed.

Before Calibration

The following should be checked and set before calibration.

(1) Contamination of the sensor

Make sure that no dirt or deposits are present on the sensor.

(2) Temperature setting

When using a sensor without a built-in temperature element, i.e., a needle type or test
tube size pH sensor, the temperature of a standard solution to be used should be set into
the meter. (Refer to Section 5.3 (2), “Manual temperature setting (M.tP) panel.”)

(3) Battery condition

Check the battery condition indicator for remaining life. If the indicator is flashing,
calibration cannot be performed. Replace the batteries. (Refer to Section 2.1, “Installing
the Batteries.”)