Yokogawa PH72 Personal pH/ORP Meter User Manual

Page 78

Advertising
background image

IM 12B03D02-01E

9-11

9. Technical Information

Obtained from equation 9.10, the potential of the Ag/AgCl electrode filled with a 3.3
mol/l KCl solution, E'

AgCl

, has the temperature characteristic for the standard hydrogen

electrode as shown in Figure 9.12.

To convert E'

AgCl

to the value for the standard hydrogen electrode,

E

: oxidation-reduction potential when reference electrode is SHE

E

: oxidation-reduction potential when reference electrode is Ag/AgCl

electrode filled with 3.3 mol/l KCl solution

E

: potential of Ag/AgCl electrode filled with 3.3 mol/l KCl solution (vs SHE)

E

: liquid junction potential (average liquid junction potential between KCl

and test solution is approximately 3 mV)

E

؍ E ؉ E ؉ E [mV]

SHE

SHE

j

(9.11)

'

AgCl

'

AgCl

j

3.3mol/l KCl-Ag/AgCl Electrode P

otential

(vs SHE), E' [mV]

60

40

20

0

Temperature [

؇

C]

150

200

250

F0912.EPS

AgCl

Figure 9.12

Temperature Characteristic of Potential of Ag/AgCl
Electrode with 3.3 mol/l KCl, E'

AgCl

(vs SHE)