9. Technical Information
Obtained from equation 9.10, the potential of the Ag/AgCl electrode filled with a 3.3
mol/l KCl solution, E'
AgCl
, has the temperature characteristic for the standard hydrogen
electrode as shown in Figure 9.12.
To convert E'
AgCl
to the value for the standard hydrogen electrode,
E
: oxidation-reduction potential when reference electrode is SHE
E
: oxidation-reduction potential when reference electrode is Ag/AgCl
electrode filled with 3.3 mol/l KCl solution
E
: potential of Ag/AgCl electrode filled with 3.3 mol/l KCl solution (vs SHE)
E
: liquid junction potential (average liquid junction potential between KCl
and test solution is approximately 3 mV)
E
؍ E ؉ E ؉ E [mV]