Orp meter, 1 orp measurement, Orp meter -1 – Yokogawa PH72 Personal pH/ORP Meter User Manual

Page 64: 1 orp measurement -1

Advertising
background image

IM 12B03D02-01E

8-1

8. ORP Meter

8. ORP Meter

8.1 ORP Measurement

Use a dedicated ORP sensor for ORP (oxidation-reduction potential) measurement. The
ORP sensor is a KCl refillable type with a platinum sensing electrode and looks the same
as a KCl refillable type pH sensor. The measuring temperature range of the ORP sensor
is 0 to 80

؇C, the same as that of the pH sensor.

• The default setting for measurement unit of the Model PH72 Personal pH/ORP

Meter is pH. Press

F/ENT

key and follow the procedure described in Section 5.3

(3), “Set measurement unit (PV.U) panel” to move to the display with mV unit.

F0801.EPS

Figure 8.1

Example of Displayed ORP Value

• Immerse the sensor in a sample solution and read after the reading has stabilized.

• To hold a measured value, press

HOLD

key. Press

DATA

key to save a

measured value. (See Section 3.4.)

Note:

Unlike pH sensors, ORP sensors do not require calibration.

A temperature element incorporated in an ORP sensor is used for measuring the
temperature of a solution, e.g., measuring the temperature of a check solution during
sensor check (see Section 8.3).

Advertising