7. Troubleshooting

7.2 Error Messages, Possible Causes, and Corrective Actions

Table 7.1

Error Message

T0701.EPS

Error Message*

1

Err1

Err2

Err3

Err4

Err5

Err6

Description

Unstable input emf

Abnormal asymmetry potential

Abnormal slope or calibration temperature

Out of measuring range

Out of temperature measuring range

Meter electronics failure

Occurrence

During calibration

During measurement

*1: For displayed digital characters, see Alphanumeric Display Table in Preface.

(1) Err1

Unstable input emf

Appears during calibration.

A calibration result is accepted when the variation of input emf for 10 seconds falls
within

؎1 mV (approximately ؎0.02 pH equivalent). If the variation for 10 seconds

cannot stabilize and is outside the range of

؎1 mV even after 3 minutes has elapsed in

calibration, an Err1 message will appear.

Possible Causes:

• Dry electrode.

• Clogged liquid junction.

• Insulation failure in electronics.

• Sensor immersed improperly.

Corrective Actions:

• Leave the sensor in a standard solution until the reading stabilizes, and then try

recalibration.

• Remove dirt or deposits from the liquid junction. (See Section 6.2.)

• Remove stain or moisture from the connectors. (See Section 6.6.)

• Immerse the sensor properly (See Section 3.2.)

• Immerse the sensor in water for 1 to 2 hours or longer to rehydrate. (See Section

6.4.)