Yokogawa PH72 Personal pH/ORP Meter User Manual

Page 32

Advertising
background image

IM 12B03D02-01E

4-5

4. Calibration

Example 2: Calibration of a sensor without a built-in temperature element* using a pH

7 standard solution

* Needle type or test tube size pH sensor

F0402.EPS

POWER

CAL

Input of standard solution

temperature

25.0

؇

C fixed

Temperature set on 'M.tP' panel

F/ENT

F/ENT

Auto calibration in progress

Auto calibration is complete

Standard
solution
temperature

to enter
standard
solution
temperature

Std. sol.

Water

Wait until
reading
stabilizes.

Wipe off washing water
from sensor thoroughly
and then immerse sensor
in standard solution.

to select auto calibration (AUto)

After
a few
seconds

Thermometer

Wash sensor with water
thoroughly (or wash in
water in beaker).

For 2-point calibration, continue the procedure in the same way as Example 1. The
difference between Example 1 and 2 is that

appears on the display and the

temperature of a standard solution should be entered manually.