2 specifications, 2 specifications -2 – Yokogawa PH72 Personal pH/ORP Meter User Manual

Page 9

Advertising
background image

IM 12B03D02-01E

1-2

1. Outline

1.2 Specifications

Measuring range:

pH:

0 to 14 pH*

1

ORP:

-2000 to 2000 mV

Temperature:

0 to 100

؇C*

2

Resolution:

pH:

0.01 pH

ORP:

0.1 mV (-199.9 to 199.9 mV)
1 mV (-2000 to -200 mV and 200 to 2000 mV)

Temperature:

0.1

؇C

Repeatability (without sensor):

pH:

؎0.01 pH ؎1 digit

ORP:

؎1 mV ؎1 digit

Accuracy (Temperature):

؎0.7؇C (0 to 70؇C)

؎1؇C (above 70؇C)

Temperature compensation (glass electrode emf — temperature characteristics):

Automatic compensation (or manual compensation*

1

)

Solution temperature:

0 to 80

؇C (or 0 to 100؇C*

3

) (0 to 50

؇C when a KCl replenish-

free type sensor and its sensor cable are immersed in water)

Solution conductivity:

50

␮S/cm or more

Ambient temperature:

0 to 50

؇C

Construction:

Protection class IP67 (IEC 60529)

Power source:

2x AA batteries (LR6)
Auto power off function (time configurable: 1 to 120 minutes)

Battery life:

Approximately 600 hours*

4

of continuous use (battery type and

operating condition dependent)

Dimensions:

Approximately 150(H) x 61(W) x 42(D) mm (not including
connector part)

Weight:

Approximately 220 g (including batteries)

EMC Conformity Standards:
EMI (Emission)

EN 61326-1 Class B

Test Item

Electromagnetic radiation disturbance

Frequency Range

30 to 1000 MHz

Basic Standard

CISPR 16-1 and 16-2

T0101.EPS

EMS (Immunity)

EN 61326-1 Table 2 (For use in inductrial locations)

Test Item

No.

1

Electrostatic
discharge

Test Specification

Basic Standard

IEC 61000-4-2

Performance

Criteria*

* A: Normal performance within the specification limits:

±

20% of the measured value.

B: Temporary degration or less of function or performance which is self-recoverable.
** Display value may be affected by strong electromagnetic field.

A

4 kV (contact)
8 kV (air)

A

2.0 to 4.0 GHz,
3 V/m (unmodulated)
80% AM (1 kHz)

2

RF amplitude
modulated
electromagnetic field

IEC 61000-4-3

B**

80 to 1000 MHz,
10 V/m (unmodulated)
80% AM (1 kHz)

A

1.4 to 2.0 GHz,
10 V/m (unmodulated)
80% AM (1 kHz)

T0102.EPS

*1: Display range is from -2 to 10 pH.
*2: Display range is from-10 to 120

؇C.

*3: When a needle type or test tube size pH sensor is used.
*4: When alkaline batteries are used.