5 the alkaline error, 5 the alkaline error -6 – Yokogawa PH72 Personal pH/ORP Meter User Manual

Page 73

Advertising
background image

IM 12B03D02-01E

9-6

9. Technical Information

9.5 The Alkaline Error

As shown in Figure 9.5, the emf of a glass electrode deviates from the linear value on
the alkaline side. This is called the alkaline error. The magnitude of the alkaline error
varies depending on the glass membrane compositions. The alkaline error is likely to
occur with the presence of sodium and lithium and even with the same pH, it varies
depending on the types and concentrations of cations and on the temperature.

Alkaline error

Acid error

EMF at pH 7

Theoretical value

Slope difference

0

2

E

(mV)

1

4

3

5

7

9

8

11

6

10

12 pH

pHx

F0905.EPS

Figure 9.5

EMF Characteristics of Glass Electrode

Test solution pH

9

0

20

40

60

A

lka

line e

rr

o

r

(mV)

10

11

13

14

F0906.EPS

12

Solution temperature
at 25

؇

C

3mol/l-Na

+

3mol/l-K

+

3mol/l-NH

4

+

Figure 9.6

Ion Type vs Alkaline Error