4. Calibration

4.1 Automatic Calibration

In automatic calibration the Model PH72 Personal pH/ORP Meter automatically
recognizes standard solutions being used and calibrates itself using values of Table 4.1.
Two types of standard solutions are preprogrammed: NIST (solutions prepared in
accordance with Japanese standards, factory default) and US (solutions prepared in
accordance with the U.S. standards). If NIST is selected, the meter recognizes standard
solutions with pH 2, 4, 7, 9, and 12. The meter recognizes standard solutions with pH 4,
7, and 10 if US is selected.

Table 4.1

pH-Temperature Relationship in pH Standard Solutions

0

5

10

15

20

25

30

35

40

45

50

pH2
pH4
pH7
pH9

pH12

-

4.003

6.984

9.464

13.423

1.668

3.999

6.951

9.395

13.207

1.670

3.998

6.923

9.332

13.003

1.672

3.999

6.900

9.276

12.810

1.675

4.002

6.881

9.225

12.627

1.679

4.008

6.865

9.180

12.454

1.683

4.015

6.856

9.139

12.289

1.688

4.024

6.844

9.102

12.133

1.694

4.035

6.838

9.068

11.984

1.700

4.047

6.834

9.038

11.841

1.707

4.060

6.833

9.011

11.705

55

60

70

80

1.715

4.075

6.834

8.985

11.574

1.723

4.091

6.836

8.962

11.449

1.743

4.126

6.845

8.921

-

1.766

4.164

6.859

8.885

-

Std. Solution

NIST

Temperature

؇

C

0

5

10

15

20

25

30

35

40

45

50

pH4
pH7

pH10

4.000

7.120

10.317

3.998

7.090

10.245

3.997

7.060

10.179

3.998

7.040

10.118

4.001

7.020

10.062

4.005

7.000

10.012

4.010

6.990

9.966

4.018

6.980

9.926

4.027

6.980

9.889

4.038

6.978

9.856

4.050

6.970

9.828

55

60

4.064

6.980

9.828

4.080

6.980

9.828

Std. Solution

Temperature

؇

C

US

T0402.EPS

Before Automatic Calibration

Make sure that the correct type of standard solutions to be used for automatic calibration
has been selected. (Refer to Section 5.3 (10), “Standard solution setting (Std) panel.”)