(2) Manual temperature setting (M.tP) panel

Used to input the temperature of a solution into the meter when using a sensor without a
built-in temperature element (needle type or test tube size pH sensor). This setting is not
required when a sensor connected has a built-in temperature element. The setting range is
from -10.0 to 110.0

؇C.

F050302.EPS

to set

to confirm

F/ENT

F/ENT

(3) Set measurement unit (PV.U) panel

Used to set a pH unit for pH measurement or a mV unit for ORP measurement.

F050303.EPS

to toggle

to
proceed

F/ENT

to confirm

F/ENT

(4) Delete all stored data (dEL.A) panel

Used to delete all stored data. Press

F/ENT

key on the flashing “dEL.A” panel.

will be flashing. Use

keys to select

. Press

F/ENT

key to delete all stored

data completely.