6.5 Replenishing the Electrode with Filling Solution (KCl

solution)

Replenishment is required only when a KCl refillable type combination sensor is used.
An electrode filling solution leaks from the liquid junction little by little during
measurement. When the level of filling solution drops to the level shown in Figure 6.3,
replenish with a 3.3 mol/l KCl solution supplied.

CAUTION

A syringe tip is sharp. Handle with care.

F0603.EPS

How to fill using syringe

Fill hole for KCl solution

Syringe (supplied)

Filling solution level

Protective cover

Replenish when filling solution
level drops to the level shown above.

2

0

mm

Figure 6.3

Replenishment of Filling Solution