Procedure – Yokogawa PH72 Personal pH/ORP Meter User Manual

Page 31

Advertising
background image

IM 12B03D02-01E

4-4

4. Calibration

Procedure

Example 1: 2-point calibration using pH 7 and pH 4 standard solutions

F0401.EPS

POWER

CAL

CAL

Auto calibration in progress

(1st point)

Auto calibration at 1st point

is complete

F/ENT

Auto calibration in progress

(2nd point)

Auto calibration at 2nd point

is complete

Calibration
at 2nd point

F/ENT

Std. sol.

Water

Wash sensor with water
thoroughly (or wash in
water in beaker).

Wait until
reading
stabilizes.

Wipe off washing water
from sensor thoroughly
and then immerse sensor
in standard solution.

Std. sol.

Water

Wash sensor with
water thoroughly.

Wait until
reading
stabilizes.

Wipe off washing water
from sensor thoroughly
and then immerse sensor
in standard solution.

to select auto calibration (AUto)

After
a few
seconds

After
a few
seconds

( to select auto calibration (AUto))

The last two calibration results are saved in the meter. Therefore, for 1-point calibration,
perform a 1-point calibration twice using the same standard solution, or initialize
calibration parameters (refer to Section 5.3 (11), “Initialize calibration parameters (I.CP)
panel”) before performing a 1-point calibration.