7 storage and o-ring/gasket replacement, 7 storage and o-ring/gasket replacement -6, Caution – Yokogawa PH72 Personal pH/ORP Meter User Manual

Page 55

Advertising
background image

IM 12B03D02-01E

6-6

6. Maintenance

6.7 Storage and O-ring/Gasket Replacement

• Storage Precautions

Care is required when storing the meter and sensor. To maintain in good condition,
observe the following:

(1) Before storage, wash off remaining sample solution from the sensor with water.

Deposits on the liquid junction, if any, must be removed thoroughly, or the
junction may be clogged. A clogged junction will cause invalid measurements,
e.g., unstable pH readings.

Keep a wetting cap attached to the sensor. For KCl refillable type sensors, seal the
fill hole with a plug.

(2) Leave the sensor connected to the meter body to protect the connectors and O-ring

from staining. Contamination may cause deteriorated insulation of connectors or
poor water resistance by the O-ring.

(3) Do not place any object on top of the sensor or on the top of the PH72 meter.

F0604.EPS

Wetting cap

Protective cover

Attach wetting cap containing cotton
wad moistened with a few drops of
water.

For KCl refillable type sensor,
seal fill hole with plug.

Figure 6.4

Sensor Storage Condition

• Storage Location

When not in use, store the meter and sensor in a safe place. If it is to be stored for a long
period, store it in a place:

• With low humidity at or near normal temperatures

• Not exposed to direct sunlight or water

• No corrosive gases are present

• Replacing the O-ring and Gasket

A gasket in the battery box and an O-ring on the sensor connector of the meter can be
replaced. Heavily contaminated or damaged gasket and/or O-ring should be replaced.

CAUTION

When installing an O-ring and gasket, clean them and their mounting surfaces with a
cloth moistened with alcohol so that they are free from dirt. Otherwise, water resistance
may not be assured.