5v input – Yokogawa µR20000 User Manual

Page 107

Advertising
background image

IM 04P02B01-01E

5-6

1-5V Input

Select 1-5V.

Set the channel range.

Set the left span value.

Set the right span value.

Set the decimal position and left scale

value.

Set the right scale value.

Turn On or Off the low-cut function.

The new setting takes effect.

(Select the right scale value.)

ESC/?

Set=Range

CH=01-01

Mode=1-5V

Span left= 1,000

Span right= 5,000

Scale left= 0.00

Scale right= 200.00

Low-cut=Off

01-01 Channel

Setting complete

First channel

Last channel

Descrpton

• Span Left and Span Rght

Selectable span left range: 0.8 to 1.2 V
Selectable span right range: 4.8 to 5.2 V

• Scale Left, Scale Rght, and Decmal Poston

Selectable range (mantissa): –20000 to 30000
Example: The value in the range of –100.00 to 350.00 cannot be specified. The

mantissa of scale right is 35000, which exceeds the upper limit of 30000.
Set in the range of –100.0 to 350.0.

Decimal position: XXXXX, XXXX.X, XXX.XX, XX.XXX, X.XXXX
The decimal position is the position specified by scale left.
Example: Change 0.00 to 0.0.
Press the

key to move the cursor to the last digit of 0.00.

Space

Press the key to change the value to 0.0 .

Make sure that scale left is less than scale right.

• Low-cut

On: Cuts values below 0% of the recording span (set to 0% (scale left value)).
Off: Does not apply the low-cut function.

<Related Topcs> Setting the unit: Section 5.3

Enabling the low-cut function: Section 7.15

5.1 Settng the Input Range

Advertising