Standard performance – Yokogawa µR20000 User Manual

Page 248

Advertising
background image

IM 04P02B01-01E

12-20

Standard Performance

Item

Specifications

Measurement and recording accuracy

The following specifications apply to operation of the recorder under standard operation conditions:

Temperature:

23 ± 2°C

Humidity:

55% ± 10%RH

Power supply voltage:

90 to 132 or 180 to 264 VAC

Power supply frequency:

50/60 Hz ± 1%

Warm-up time:

At least 30 minutes.

Other ambient conditions such as vibration should not adversely affect the operation

Input Type Range Type

Measurement (Digital Display)

Recording (Trend)

Measurement Accuracy

Highest Res. Recording Accuracy

Resolution

DC voltage

20 mV

±(0.1% of rdg + 2 digits)

10 μV

Measurement

accuracy

±(0.3% of recording

span)

Pen model:

Dead band 0.2%

of recording span
Dot model:

Resolution 0.1 mm

60 mV

10 μV

200 mV

100 μV

2 V

1 mV

6 V

1 mV

20 V

10 mV

50 V

±(0.1% of rdg + 2 digits)

10 mV

DC voltage

1-5 V

±(0.1% of rdg + 2 digits)

1 mV

TC
(excludes

RJC

accuracy)

R

S

B

±(0.15% of rdg + 1)

except R,S: 0 to 100°C, ±3.7°C

100 to 300°C, ±1.5°C

B: 400 to 600°C, ±2°C

Accuracy not warranted for

values less than 400°C

0.1°C

K

±(0.15% of rdg + 0.7°C)

except ±(0.15% of rdg + 1°C)

for –200 to –100°C

E

J

T

±(0.15% of rdg + 0.5°C)

except ±(0.15% of rdg + 0.7°C)

for J: –200 to –100°C

N

±(0.15% of rdg + 0.7°C)

W

±(0.15% of rdg + 1°C)

L

U

±(0.15% of rdg + 0.5°C)

except ±(0.15% of rdg + 0.7°C)

for L:–200 to –100°C

WRe

±(0.2% of rdg + 1°C)

RTD

Pt100

JPt100

±(0.15% of rdg + 0.3°C)

ON/OFF input Level

Contact

Threshold level (2.4 V)

accuracy ± 0.1 V

≤ 1 kΩ: ON, ≥ 100 kΩ: OFF

(Parallel capacitance

0.01 μF or less)

Measuring accuracy in case of linear scaling

Accuracy during scaling (digits) = measuring accuracy (digits) × multiplier + 2 digits (rounded up)

where the multiplier = scaling span (digits)/measuring span (digits).

Example: Assuming that

• range type:

1-5 V

• measuring span:

1.000 to 5.000 V

• scaling span:

0.000 to 2.000 Then,

Measuring accuracy = ±(0.1% × 5 V + 2 digits)

= ±(0.005 V [5 digits] + 2 digits)

= ±7 digits

Multiplier = {2000 digits (0.000 to 2.000)}/4000 digits (1.000 to 5.000) = 0.5

Accuracy during scaling = ±(7 × 0.5 + 2) digits = 6 digits (rounded up)

12.6 General Specifications