Yokogawa µR20000 User Manual

Page 218

Advertising
background image

IM 04P02B01-01E

10-6

Is the input free of noise?

NO

YES

• Separate the input and

communication wiring from the

noise source.

• Ground the recorder properly.

• Use shielded cables for input and

communication wiring.

• Insert an isolation transformer

into the power line.

• Insert a power line filter into the

power line.

• If an inductive load is connected

to an alarm contact output, use a

surge suppresser on that line.

Contact your nearest
YOKOGAWA dealer.

Display and other functions do not work properly.

Is the key lock released?

NO

YES

Release the key lock using the
FUNC key (see section 3.13).

Contact your nearest
YOKOGAWA dealer.

Keys do not work.

Is the customize

menu released?

NO

YES

Disable the Customized Menu
with a key operation (see section
7.22).

Contact your nearest
YOKOGAWA dealer.

Some of the menus are not displayed.
Unable enter Basic Setting mode.

10.2 Troubleshootng Flow Charts