9-29

IM 04P02B01-01E

Operat

ons
Related

to

the
Computat

on

Funct

on

(/M1
Opt
on)

1

2

3

4

5

6

7

8

9

10

11

12

App

Index

Explanaton

Error

Specifies how to display or print the computed result when computation errors occur.
+Over: Set to +over. Displayed/Printed as “+Over.”
–Over: Set to –over. Displayed/Printed as “–Over.”

Computation that result in errors
• x/0
• SQR(-x)
• LOG(-x)
• When a channel set to Skip is included in the equation

Over

Selects the procedure when an “over” value

*

is input for TLOG.SUM or TLOG.AVE

computation.

* An “over” value on a measurement channel refers to +Over or –Over. See section 12.4. An

“over” value on a computation channel refers to a value that is less than or equal to
–100000000 or greater than or equal to 100000000 excluding the decimal.

Skip: The “over” value is not used in the computation.
Limit: The following limit value is used for the computation.

Channel

Lmt Value

Measurement channel

• For scaling (1-5V, scaling, and square root)
–Over: –315000, +Over: 31500 (excluding the decimal)
• For input types other than scaling
–Over: The minimum measurable value at the specified range type

(Example: –2.000 at 2 V range)

+Over: The maximum measurable value at the specified range type

(Example: 2.000 at 2 V range)

Computation channel

–Over: –100000000, +Over: 100000000 (excluding the decimal)

9.17 Settng the Procedure Taken When the Computed Result Is n Error