background image

5-7

IM 04P02B01-01E

Frequently

Used

Setup

Operat

ons
(Sett

ng

Mode)

1

2

3

4

5

6

7

8

9

10

11

12

App

Index

Note

• The displayable/printable range of scale values is –19999 to 30000 excluding the decimal.
• If the leftmost value of the scale is set to –20000 (excluding the decimal), the value when

the low-cut function is applied is displayed as “–Over.”

Square Root Computaton

Select SQRT.

Set the channel range.

Select the range type of DC voltage.

20 mV, 60 mV, 200 mV, 2 V, 6 V, 20 V, or 50 V

Set the left span value.

Set the right span value.

Set the decimal position and

left scale value.

Set the right scale value.

Turn On or Off the low-cut function.

(When low-cut is set to Use)

(When low-cut is On)

Set the low-cut point.

0.0 to 5.0% of the recording span, 0.1 steps.

The new setting takes effect.

ESC/?

Set=Range

CH=01-01

Mode=SQRT

Range=2V

Span left= -2.000

Span right= 2.000

Scale left= 0.00

Scale right= 200.00

Low-cut=Off

Low-cut point=0.0

01-01 Channel

Setting complete

First channel

Last channel

Descrpton

• Input Type n Whch Square Root Computaton Can Be Performed

Square root computation can be performed when set to DC voltage input.

• Span Left and Span Rght

Same as the selectable span range of the DC voltage input.
Make sure that span left is less than span right.

5.1 Settng the Input Range