Channel<x>:offset, Channel<x>:probe – GW Instek GBS-1000 Series Programming Manual User Manual

Page 28

Advertising
background image

GBS-1000 Programming Manual

28

:CHANnel<X>:OFFSet

Set

Query

Description

Sets or returns the offset level for each channel. The
offset level range is dependent on the vertical scale.

Syntax

< Long >

< Short >

:channel<X>:offset <NR3>
:channel<X>:offset?

:chan<X>:offs <NR3>
:chan<X>:offs?

Parameter

<X>

Channel

<NR3> Offset level

1/2/3/4 CH1/2/3/

4

±0.5

–0.5V ~ +0.5V
(2mV/div~50mV/div)

±5.0

–5.0V ~ +5.0V
(100mV/div~500mV/div)

±50.0 –50.0V ~ +50.0V

(1V/div ~ 5V/div)

Example

:channel1:scale 1.00e–2
:channel1:offset 2.00e–2

Set CH1 scale to 10mV
Set CH1 offset to 20mV

:CHANnel<X>:PROBe

Set

Query

Description

Sets or returns the probe attenuation factor.
Same as: Channel key → F4

Syntax

< Long >

< Short >

:channel<X>:probe <NR3>
:channel<X>:probe?

:chan<X>:prob <NR1>
:chan<X>:prob?

Parameter

<X>

Channel

<NR1> Probe attenuation factor

1/2/3/4 CH1/2/3/4 0

1x

1

10x

2

100x

Example

:channel1:probe 1

Set CH1 probe to 10x

Advertising