GW Instek GSP-930 User Manual User Manual
Page 149

ADVANCED OPERATION
147
Rel Stop[F5]
Sets the relative stop
level for the selected
offset.
Man: Allows a user-
defined Abs Stop level.
Couple: Sets the Rel Stop
level to the Rel Start
level.
7. Press Fail Mask[F6] to set the Fail Mask
conditions:
Absolute[F1]
Sets the fail condition to
the Absolute level limit.
Relative[F2]
Sets the fail condition to
the relative level limit.
Abs AND Rel[F3]
Sets the fail condition as
both the absolute and
relative level limits.
Abs OR Rel[F4]
Sets the fail condition to
either the absolute or
relative level limits.
8. Press Select Offset[F1] and repeat the above
steps for any other offsets.
Offset:
1~5
Pre-Set Test
Parameters:
3GPP
For details on 3GPP SEM test parameters,
please see the SEM overview on page 132.
1. Press Setup[F1]>3GPP[F1] to choose 3GPP
measurement.