Single sweep – GW Instek GSP-930 User Manual User Manual
Page 69

BASIC OPERATION
67
Operation
1. Press
Sweep
> Sweep Time[F1] and toggle the
Sweep time to Auto or Man.
2. Set the sweep time for the Man mode.
Mode:
Range:
Resolution:
Auto, Man
22ms ~ 1000s (span>0Hz)
50us ~ 1000s (span=0Hz)
10us
Display Icon
The SWT icon will be displayed at the
bottom of the screen when in the sweep is
in manual mode.
Single Sweep
Description
The single sweep function is used to perform a
single sweep. When Sweep Single is pressed the
GSP-930 will perform a single sweep and then
stop.
Operation
1. Press
Sweep
> Sweep Single[F2] to put the
spectrum analyzer into single sweep mode.
2. Press Sweep Single[F2] again to perform a single
sweep.
Display Icon
The Sweep Single icon is displayed on the
right-hand side of the screen when the
sweep is in single mode.
Note
You must wait for the single sweep to finish before
pressing the Single Sweep key again.
If a setting is changed whilst the spectrum analyzer
is still sweeping, the single sweep will immediately
start over.