GSP-930 User Manual
54
Correction Table
5. Repeat steps 2 to 4 for any other correction
points.
6. To delete the selected point, press Delete
Point[F6].
7. Press Return[F7]>Save Correction[F5] to save the
correction set.
Note
Note that the correction points are automatically
sorted by frequency (low → high).
The correction set must be saved before it can be
turned on.
The frequency values
displayed
in the correction
table are rounded down for display purposes only.
The actual frequency for each point can be seen in
the Frequency soft-key.
Amplitude Correction On/Off
Description
Any one of the 5 correction sets can be turned
on.
Activate
Correction
1. Press
Amplitude
> More[F7]>Correction[F3]>
Correction Set[F1] and choose a correction set.
Correction Set:
1~5
2. Press Correction[F2] and toggle correction on.