GW Instek GSP-930 User Manual User Manual
Page 166

GSP-930 User Manual
164
Operation
1. Press
Limit
Line
>Pass/Fail Test.
2. To set a high limit, press High Limit[F1] and
choose one of the limit lines as the upper (high)
limit.
3. To set the low limit, press Low Limit[F2] and
select one of the limit lines as the lower limit.
4. Press Pass Criterion[F3] and select the pass
criteria.
Criteria:
All-In, Max-In, Min-In
5. Press Pass/Fail Test and turn the testing on.
6. The test result appears in the bottom of the
display.
Pass:
Fail:
Display Icon
The alarm icon is shown at the bottom of
the display whenever testing is turned on.
Note
At least one limit line (high or low) must be turned
on to enable testing.
If the high limit or low limit is turned off, the
maximum or minimum* display level is set
automatically as the high or low limit, respectively.
* +30dBm+Ref level offset or -120dBm+Ref level
offset