Trace detection mode – GW Instek GSP-930 User Manual User Manual
Page 77
BASIC OPERATION
75
4. Select the trace math function:
PowerDiff[F3]
LogDiff[F4]
LogOffset[F5]
5. If LogDiff was selected, set the reference level
and unit.
LogDiff ref range:
LogDiff ref units:
-120dBm ~ 30dBm
dBm, W
6. If LogOffset was selected, set the offset level
and unit.
LogOffset range:
-50dB~+50dB
7. To turn trace math off, press the OFF[F6].
Display Icon
The Math icon is displayed when trace
math is turned on.
Trace Detection Mode
Description
Each time the spectrum analyzer samples data
for each point on the trace, a number of samples
are usually taken for each point, known as a
sample bucket. The actual value of each point is
determined by the detector from the samples in
each bucket.
Each selected trace, (1, 2, 3, 4), can use a
different detection mode.