System:version – GW Instek AFG-2225 User Manual User Manual
Page 194

AFG-2225 Series User Manual
194
*RST
System Command
Description
Reset the function generator to its factory default
state.
Note
Note the *RST command will not delete
instrument save states in memory.
Syntax
*RST
SYSTem:VERSion?
System Query
Description
Performs a system version query. Returns a string
with the instrument, firmware version, FPGA
revision
Query Syntax
SYSTem:VERSion?
Return parameter <string>
Example
SYST:VERS?
AFG-2225 VX.XXX_XXXX
Returns the year and version for that year (1).
*OPC
System Command
Description
This command sets the Operation Complete Bit
(bit 0) of the Standard Event Status Register after
the function generator has completed all pending
operations. For the AFG-2225, the *OPC command
is used to indicate when a sweep or burst has
completed.
Note
Before the OPC bit is set, other commands may be
executed.
Syntax
*OPC