Source[1|2]:sweep:source – GW Instek AFG-2225 User Manual User Manual
Page 246

AFG-2225 Series User Manual
246
Example
SOUR1:SWE:TIME?
+2.0000E+01
Returns the sweep time (20 seconds).
SOURce[1|2]:SWEep:SOURce
Source Specific
Command
Description
Sets or queries the trigger source as immediate
(internal), external or manual. Immediate (internal)
is the default trigger source. IMMediate will
constantly output a swept waveform. EXTernal
will output a swept waveform after each external
trigger pulse. Manual will ouput a swept
waveform after the trigger softkey is pressed.
Note
If the APPLy command was used to create the
waveform shape, the source is automatically set to
IMMediate.
The *OPC/*OPC? command/query can be used to
signal the end of the sweep.
Syntax
SOURce[1|2]: SWEep:SOURce {IMMediate|EXTernal|
MANual}
Example
SOUR1: SWE:SOUR EXT
Sets the sweep source to external.
Query Syntax
SOURce[1|2]: SWEep:SOURce?
Return Parameter IMM
Immediate
EXT
External
MANual
Manual
Example
SOUR1:SWE:SOUR?
IMM
The sweep source is set to immediate.