Source[1|2]:sum:state, Source[1|2]:sum:source – GW Instek AFG-2225 User Manual User Manual
Page 237

REMOTE INTERFACE
237
SOURce[1|2]:SUM:STATe
Source Specific
Command
Description
Turns SUM Modulation on or off. By default SUM
modulation is off.
Note
Burst or sweep mode will be disabled if SUM
modulation is enabled. As only one modulation is
allowed at any one time, other modulation modes
will be disabled when SUM modulation is
enabled.
Syntax
SOURce[1|2]:SUM:STATe {OFF|ON}
Example
SOUR1:SUM:STAT ON
Enables SUM modulation
Query Syntax
SOURce[1|2]:SUM:STATe?
Return Parameter 0
Disabled (OFF)
1
Enabled (ON)
Example
SOUR1:SUM:STAT?
ON
SUM modulation is currently enabled.
SOURce[1|2]:SUM:SOURce
Source Specific
Command
Description
Sets or queries the SUM source as internal or
external. Internal is the default source.
Note
If an external SUM source is selected, the
amplitude is controlled by the MOD INPUT
terminal on the rear panel.
Syntax
SOURce[1|2]:SUM:SOURce {INTernal|EXTernal}
Example
SOUR1:SUM:SOUR EXT
Sets the SUM source to external.
Query Syntax
SOURce[1|2]:SUM:SOURce?
Return Parameter INT
Internal