Source[1|2]:pm:internal:function, Source[1|2]:pm:internal:frequency – GW Instek AFG-2225 User Manual User Manual
Page 234

AFG-2225 Series User Manual
234
Example
SOUR1:PM:SOUR?
INT
The PM source is set to internal.
SOURce[1|2]:PM:INTernal:FUNction
Source Specific
Command
Description
Sets the shape of the modulating waveform from
sine, square, triangle, upramp and dnramp. The
default shape is sine.
Note
Square and triangle waveforms have a 50% duty
cycle. Upramp and dnramp have a symmetry to
100% and 0%, respectively. .
Syntax
SOURce[1|2]:PM:INTernal:FUNction
{SINusoid|SQUare|TRIangle|UPRamp|DNRamp}
Example
SOUR1:PM:INT:FUN SIN
Sets the PM modulating wave shape to sine. .
Query Syntax
SOURce[1|2]:PM:INTernal:FUNction?
Return Parameter SIN
Sine
UPRAMP
Upramp
SQU
Square
DNRAMP
Dnramp
TRI
Triangle
Example
SOUR1:PM:INT:FUNC?
SIN
The shape for the modulating waveform is Sine.
SOURce[1|2]:PM:INTernal:FREQuency
Source Specific
Command
Description
Sets the modulating waveform frequency for
internal sources. The default frequency is set to
100Hz.
Syntax
SOURce[1|2]:PM:INTernal:FREQuency
{<frequency>|MINimum|MAXimum}
Parameter
<frequency>
2 mHz~ 20 kHz