Selecting the (am) modulation source – GW Instek AFG-2225 User Manual User Manual
Page 73

MODULATION
73
Range
Depth
0%~120%
Default depth
100%
Note
When the modulation depth is greater than 100%, the
output cannot exceed ±5VPeak (10kΩ load).
If an external modulation source is selected,
modulation depth is limited to ± 5V from the MOD
INPUT terminal on the rear panel. For example, if the
modulation depth is set to 100%, then the maximum
amplitude is +5V, and the minimum amplitude is -5V.
Selecting the (AM) Modulation Source
The function generator will accept an internal or external source for
AM modulation. The default source is internal.
Panel Operation
1. Press the MOD key.
2. Press F1 (AM).
3. Press F1 (Source).
our
4. Press F1 (INT) or F2 (EXT) to
select the modulation source.
~
5. Press Return to go back to the
previous menu.
Return
External Source
Use the MOD INPUT terminal
on the rear panel when using an
external source.
IN
OUT
Trigger
Counter
Trigger
MOD