Status register commands – GW Instek AFG-3000 Series User Manual
Page 200

AFG-3000 Series User Manual
200
Status Register Commands
*CLS
System Command
Description
The *CLS command clears all the event registers,
the error queue and cancels an *OPC command.
Syntax
*CLS
*ESE
System Command
Description
The Standard Event Status Enable command
determines which events in the Standard Event
Status Event register can set the Event Summary
Bit (ESB) of the Status Byte register. Any bit
positions set to 1 enable the corresponding event.
Any enabled events set bit 5 (ESB) of the Status
Byte register.
Note
The *CLS command clears the event register, but
not the enable register.
Syntax
*ESE <enable value>
Parameter
<enable value>
0~255
Example
*ESE 20
Sets a bit weight of 20 (bits 2 and 4).
Query Syntax
*ESE?
Return Parameter Bit Register
Bit Register
0 Not
used 4 Message
Available
1 Not
used 5 Standard
Event
2 Error
Queue 6 Master
Summary
3 Questionable
Data
7 Not
used