Source[1]:burst:internal:period – GW Instek AFG-3000 Series User Manual
Page 255

REMOTE
INTERFACE
255
Example
SOUR1:BURS:NCYC?
+1.0000E+02
The burst cycles are set to 100.
SOURce[1]:BURSt:INTernal:PERiod
Source Specific
Command
Description
Sets or queries the burst period. Burst period
settings are only applicable when the trigger is set
to immediate. The default burst period is 10 ms.
During manual triggering, external triggering or
Gate burst mode, the burst period settings are
ignored.
Note
The burst period must be long enough to output
the designated number of cycles for a selected
frequency.
Burst period > burst count/(waveform frequency
+ 200 ns)
If the period is too short, it is automatically
increased so that a burst can be continuously
output. A “data out of range” error will also be
generated.
Syntax
SOURce[1]:BURSt:INTernal:PERiod
{<seconds>|MINimum|MAXimum}
Parameter
<seconds > 1 us ~ 500 seconds
Example
SOUR1:BURS:INT:PER +1.0000E+01
Sets the period to 10 seconds.
Query Syntax
SOURce[1]:BURSt:INTernal:PERiod?
[MINimum|MAXimum]
Return Parameter <NR3>
Returns the burst period in seconds.
Example
SOUR1:BURS:INT:PER?
+1.0000E+01
The burst period is 10 seconds.