Save and recall commands – GW Instek AFG-3000 Series User Manual
Page 273

REMOTE
INTERFACE
273
Outputs the current arbitrary waveform in
memory.
Save and Recall Commands
Up to 10 different instrument states can be stored to non-volatile
memory (memory locations 0~9).
*SAV
Instrument
Command
Description
Saves the current instrument state to a specified
save slot. When a state is saved, all the current
instrument settings, functions and waveforms are
also saved.
Note
The *SAV command doesn’t save waveforms in
non-volatile memory, only the instrument state.
The *RST command will not delete saved
instrument states from memory.
Syntax
*SAV {0|1|2|3|4|5|6|7|8|9}
Example
*SAV 0
Save the instrument state to memory location 0.
*RCL
Instrument
Command
Description
Recall previously saved instrument states from
memory locations 0~9.
Syntax
*RCL {0|1|2|3|4|5|6|7|8|9}
Example
*RCL 0
Recall instrument state from memory location 0.
MEMory:STATe:DELete
Instrument
Command
Description
Delete memory from a specified memory location.