Source[1]:apply:noise, Source[1]:apply:triangle – GW Instek AFG-3000 Series User Manual
Page 208

AFG-3000 Series User Manual
208
Parameter
<frequency>
500μHz~50MHz
<
amplitude
>
10mV~10V (50Ω)
<offset>
0~4.99V (50Ω)
Example
SOUR1:APPL:PULS 1KHZ,MIN,MAX
Sets frequency to 1kHz and sets the amplitude to
minimum and the and offset to the maximum.
SOURce[1]:APPLy:NOISe
Source Specific
Command
Description
Outputs Gaussian noise with a 50 MHz
bandwidth. Amplitude and offset can also be set.
Note
Frequency cannot be used with the noise function;
however a value (or DEFault) must be specified.
The frequency is remembered for the next function
used.
Syntax
SOURce[1]:APPLy:NOISe [<frequency|DEFault>
[,<amplitude> [,<offset>] ]]
Parameter
<frequency>
Not applicable
<
amplitude
>
10mV~10V (50Ω)
<offset>
0~4.99V (50Ω)
Example
SOUR1:APPL:NOIS DEF, 3.0, 1.0
Sets the amplitude to 3 volts with an offset of 1
volt.
SOURce[1]:APPLy:TRIangle
Source Specific
Command
Description
Outputs a triangle wave from the selected channel
when the command has executed. Frequency,
amplitude and offset can also be set.
Syntax
SOURce[1]:APPLy:TRIangle [<frequency>
[,<amplitude> [,<offset>] ]]
Parameter
<frequency>
1μHz~1MHz
<
amplitude
>
10mV~10V (50Ω)