Source[1]:voltage:unit – GW Instek AFG-3000 Series User Manual
Page 219

REMOTE
INTERFACE
219
the output termination from high impedance to
50Ω will half the amplitude.
If the output termination is set to high impedance,
dBm units cannot be used. The units will default to
Vpp.
Syntax
OUTPut:LOAD {DEFault|INFinity}
Example
OUTP:LOAD DEF
Sets the output termination to 50Ω.
Query Syntax
OUTPut:LOAD?
Return Parameter DEF Default
INF INFinity
Example
OUTP:LOAD?
DEF
The output is set to the default of 50Ω.
SOURce[1]:VOLTage:UNIT
Source Specific
Command
Description
Sets or queries the output amplitude units. There
are three types of units: VPP, VRMS and DBM.
Note
The units set with the VOLTage:UNIT command
will be used as the default unit for all amplitude
units unless a different unit is specifically used for
a command.
If the output termination is set to high impedance,
dBm units cannot be used. The Units will
automatically default to Vpp.
Syntax
SOURce[1]:VOLTage:UNIT {VPP|VRMS|DBM}
Example
SOUR1:VOLT:UNIT VPP
Sets the amplitude units to Vpp.
Query Syntax
SOURce[1]:VOLTage:UNIT?
Return Parameter VPP Vpp
VRMS Vrms