Caution, Pvcheck – HT instruments PVCHECK User Manual
Page 19

PVCHECK
EN - 17
5.4. M
– SETTINGS FOR INSULATION MEASUREMENT
5.4.1. Instrument
settings
1. Position the cursor onto M
by using the arrow keys
(
,) and confirm with ENTER. The display shows the
following screen:
15/05/12 15:34:26
I n s . T e s t :
1 0 0 0 V
R i m i n
:
1 . 0
M
M o d e :
F i e l d
V t e s t - - - V - - -
V
R i ( + )
- - -
M
R i ( - )
- - -
M
R p
- - -
M
Selection
M
2. Press the ENTER key. The instrument shows the option:
Settings:
3. Confirm with ENTER. The instrument shows the following
screen:
15/05/12 15:34:26
I n s . T e s t
:
1 0 0 0
V
R i m i n
:
1 . 0
M
M o d e :
F i e l d
V t e s t - - - V - - -
V
R i ( + )
- - -
M
R i ( - )
- - -
M
R p
- - -
M
Settings
Selection
M
4. Use the arrow keys (
,) for option setting, and select
the desired option by means of the arrow keys (
, ).
The following parameters can be set:
Ins. Test test voltage: 250, 500, 1000VDC
Mode operating modes: Field, Timer, String
Ri min minimum limit value for insulation resistance
Test time maximum value of test time (only for
TIMER mode) with a maximum value of 300s in steps
of 1s.
5. Press the SAVE key to save the settings made; the
message “Data saved” will be displayed for a few
seconds. Press the ESC/MENU key to exit without saving
and go back to the previous screen.
15/05/12 15:34:26
I n s . T e s t : 1 0 0 0 V
R i m i n
: 1 . 0 0 M
M o d e
: T I M E R
T e s t t i m e : 1 0 s
S A V E t o s t o r e
M
CAUTION
The settings saved for test voltage also affect the settings for the insulation
measurement contained in the IVCK measurement (MENU
IVCK)