HT instruments PVCHECK User Manual
Page 56

PVCHECK
EN - 54
7.3.2. Recalling on the display the results of IVCK, M
and LOW measurements
1. Press the ESC/MENU key to go back to the main menu,
select “MEM” and confirm with ENTER to access the
section where saved values are displayed. The screen to
the side is shown by the instrument and contains the
saved tests.
2. Using the arrow keys (
,) and the arrow key , select
“Recall”, then “IVCK” and confirm with ENTER to display
the results of the I-V curve only.
3. The field “DATE” indicates the date/time in which the
result of measurement has been saved, the field “TYPE”
indicates the type of test carried out (LOW
, M, IVCK).
15/05/12 15:34:26
MEM DATE TYPE
001
08/04/2012 10:38 LOW
002
13/04/2012 12:15 M
003
15/05/12 12:20 IVCK
E f f i c i e n c y
R e c I V C K , M
, L O W
O p e n
D e l e t e
Selection
MEM- IVCK
4. Use the arrow key
to go to the “Comments” label
5. The instrument will display the comment inserted by the
operator during the saving procedure (see § 7.2) as
regards the system.
6. The symbol “*” next to the measurement number indicates
that the instrument has performed the tests with recording
of irradiance and temperature values through the remote
unit, but these values have not been transferred or are not
available. For these measurements, the values
translated to STC will not be available.
7. Press ESC/MENU to exit the screen and go back to the
main menu.
15/06/12 15:34:26
MEM Comments
001
PV PLANT 1
002*
PV PLANT 2
Selection
MEM- IVCK