4.5.2 Manufacturer specific parameters of the signal converter for Resource Block

Manufacturer specific parameters for Resource Block

Parameter

Parameter

Parameter

Parameter
DD Name

Access

Description
Settings

Initial

DEV_DESCRIPTION

DEV_DESCRIPTION

DEV_DESCRIPTION

DEV_DESCRIPTION

device description

Describes some characteristics of the device.

SERIAL_NUMBER

Serial Number

R

device serial number , serial number of the system

current serial

number

PRODUCTION_ORDER

production order

R

Production order for complete meter

current production

order

ELECTRONIC_REV

electronic revision

R

Electronic Revision of the device

current electronic

revision

DIAGNOSIS

DIAGNOSIS

DIAGNOSIS

DIAGNOSIS

Diagnosis

detailed diagnosis of the device.

MEMORY_SUMMARY

Memory Summery

R

summary of self diagnosis.

ok

Setting:

•

ok

•

bit8…15: not used

•

EEPROM

•

FRAM

•

FLASH

•

ROM application

•

ROM boot area

•

RAM ext

•

RAM int

•

Task Control

FRAM_DIAG

FRAM diagnosis

R

Result of FRAM self diagnosis.

ok

Setting:

•

ok

•

not supported action

•

unknown

•

busy

•

wait

•

first init

•

changed

•

bit09

•

parameter error

•

page error

•

chip error

•

write error

•

warning: inconsistent

•

warning: uncertain

•

warning: write cycles

•

warning: one header

•

ok, page high

work_LOAD

work load

R

Work load of the processor.

current work load

DIAGNOSIS_BLK

DIAGNOSIS_BLK

DIAGNOSIS_BLK

DIAGNOSIS_BLK

diagnosis block