Industrial Registers, Models ER-420-AC, ER-420-DC & ER-420-LP

Display flow rate / total or flow rate
This is the main display information of the ER-420 After selecting any other information, it will always return to this main

display automatically
Total is displayed on the upper line of the display and Flow Rate on the bottom line
It is possible to display Flow Rate only with the large 0 67" digits; in this instance press, SELECT to read the Total
When "-------" is shown, then the Flow Rate value is too high to be displayed The arrows indicate the increase or decrease of

the Flow Rate trend
Clear total
The value for Total can be re-initialized by pressing CLEAR twice After pressing CLEAR once, the flashing text

"PUSH CLEAR" is displayed To avoid re-initialization at this stage, press a key other than CLEAR or wait for 20 secs
Re-initialization of the Total DOES NOT influence the Accumulated Total
Display accumulated total
When you press SELECT, the Total and Accumulated Total display The Accumulated Total cannot be re-initialized The value will

count up to 99,999,999,999 The unit and number of decimals are displayed according to the configuration settings for Total
Display multiplication factor for total and accumulated total
When you press SELECT again, the actual multiplier factor is displayed for both totalizers If, for example, the factor is “x100”

and Total displays “54321 USGAL”, then the true measured volume is 5432100 USGAL

OTEE:

N

This multiplier does NOT affect the displayed Flow Rate!

Alarm 01-03
When the alarm flag starts to blink, an internal alarm condition has occurred Press SELECT several times to display the 5-digit

error code The codes are:
• 0001: irrecoverable display-data error: data on the display might be corrupted
• 0002: irrecoverable data-storage error: the programming cycle might have gone wrong: check programmed values
• 0003: error 1 and error 2 occurred simultaneously
The alarm condition will almost certainly be handled internally and if all mentioned values still appear correct, no intervention

by the operator is needed If the alarm occurs more often or stays active for a longer time, please contact your supplier

Page 10