OUTPut:EXTernal:MODulation
Description:
Set or return the output modulation state for the 2530 external calibrator. If set to CW, a
calibrated CW signal will be generated. If set to PULSE, the output will be pulse modulated
as specified by the
OUTPut:EXTernal:PULse:PERiod
,
:DCYC
, and
:SOURce
settings.
Syntax:
OUTPut:EXTernal:MODulation <asc>
Argument:
<asc> = CW, PULSE
Valid Modes:
Any
OUTPut:EXTernal:PULse:PERiod
Description:
Set or return the pulse period for the 2530 external calibrator’s internal pulse modulator.
Syntax:
OUTPut:EXTernal:PULse:PERiod <asc>
Argument:
<asc> = 0.1, 1, 10 milliseconds
Valid Modes:
Any
OUTPut:EXTernal:PULse:DCYC
Description:
Set or return the pulse duty cycle for the 2530 external calibrator’s internal pulse modulator.
Syntax:
OUTPut:EXTernal:PULse:DCYC <asc>
Argument:
<asc> = 10, 20, 30, 40, 50, 60, 70, 80, 90 %
Valid Modes:
Any
OUTPut:EXTernal:PULse:SOURce
Description:
Set or return the pulse modulation source for the 2530 external calibrator. If set to INT, the
internal pulse generator is used. EXT selects the signal from the calibrator’s rear-panel “Ext
Pulse” input. Note this input is 50 ohms, and requires TTL compatible signal levels.
Syntax:
OUTPut:EXTernal:PULse:SOURce <asc>
Argument:
<asc> = INT, EXT
Valid Modes:
Any
OUTPut:RECOrder:SIGNal
Description:
Set or return the on/off state of the instrument’s rear-panel Recorder Output port.
Syntax:
OUTPut:RECOrder:SIGNal <asc>
Argument:
<asc> = OFF, ON
Valid Modes:
Any
OUTPut:RECOrder:SOURce
Description:
Set or return the source channel for the recorder output.
Syntax:
OUTPut:RECOrder:SOURce <asc>
Argument:
<asc> = CH1, CH2
Valid Modes:
Any