4-45

Boonton Electronics

Chapter 4

4530 Series RF Power Meter

Remote Operation

OUTPut:EXTernal:MODulation

Description:

Set or return the output modulation state for the 2530 external calibrator. If set to CW, a
calibrated CW signal will be generated. If set to PULSE, the output will be pulse modulated
as specified by the

OUTPut:EXTernal:PULse:PERiod

,

:DCYC

, and

:SOURce

settings.

Syntax:

OUTPut:EXTernal:MODulation <asc>

Argument:

<asc> = CW, PULSE

Valid Modes:

Any

OUTPut:EXTernal:PULse:PERiod

Description:

Set or return the pulse period for the 2530 external calibrator’s internal pulse modulator.

Syntax:

OUTPut:EXTernal:PULse:PERiod <asc>

Argument:

<asc> = 0.1, 1, 10 milliseconds

Valid Modes:

Any

OUTPut:EXTernal:PULse:DCYC

Description:

Set or return the pulse duty cycle for the 2530 external calibrator’s internal pulse modulator.

Syntax:

OUTPut:EXTernal:PULse:DCYC <asc>

Argument:

<asc> = 10, 20, 30, 40, 50, 60, 70, 80, 90 %

Valid Modes:

Any

OUTPut:EXTernal:PULse:SOURce

Description:

Set or return the pulse modulation source for the 2530 external calibrator. If set to INT, the
internal pulse generator is used. EXT selects the signal from the calibrator’s rear-panel “Ext
Pulse” input. Note this input is 50 ohms, and requires TTL compatible signal levels.

Syntax:

OUTPut:EXTernal:PULse:SOURce <asc>

Argument:

<asc> = INT, EXT

Valid Modes:

Any

OUTPut:RECOrder:SIGNal

Description:

Set or return the on/off state of the instrument’s rear-panel Recorder Output port.

Syntax:

OUTPut:RECOrder:SIGNal <asc>

Argument:

<asc> = OFF, ON

Valid Modes:

Any

OUTPut:RECOrder:SOURce

Description:

Set or return the source channel for the recorder output.

Syntax:

OUTPut:RECOrder:SOURce <asc>

Argument:

<asc> = CH1, CH2

Valid Modes:

Any