7 interface, 8 environmental specifications – Boonton 4530 Peak Power Meter User Manual User Manual

Page 19

Advertising
background image

Boonton Electronics

Chapter 1

4530 Series RF Power Meter

General Information

1-5

1.4.7 Interface

Video Output:

Compressed representation of detected RF envelope of peak channel(s)
envelope for external oscilloscope monitor or external device synchroni-
zation. This output is roughly logarithmic with input power, is not
calibrated, and can not be used for making any measurements.

Recorder Output:

Programmable voltage output which may be used for monitoring mea-
surements or status of either channel, or for outputting a fixed,
programmable voltage. When used as a measurement monitor, the out-
put is proportial to displayed signal level with programmable or automatic
scaling.

Output range:

0 to 10V (unipolar), or -10V to +10V (bipolar)

Output resolution:

5.0mV

Output impedance:

9K

Absolute accuracy:

±100mV typical, ±200mV max, uncalibrated

±20mV after user calibration

Linearity:

0.1% typical

GPIB Interface:

Complies with IEEE-488.1. Implements AH1, SH1,T6, LE0, SR1, RL1, PP0,
DC1, DT1, C0, and E1.

RS-232 Interface:

Accepts GPIB commands (except bus dependent commands). Provides
for user software updates.

Remote Programming:

SCPI (1990) compliant and Native Mode commands via GPIB or RS-232
interfaces.

Software Drivers:

LABVIEW drivers available.

1.4.8 Environmental Specifications

General:

Manufactured to the intent of MIL-T 28800E, Type III, Class 5, Style E

CE Mark:

Conforms to EU specifications:

EN 61010-1(90)(+A1/92)(+A2/95)
EN 61010-2-031
EN 61326-1(97)
EN 55022(94)(A2/97)Class B

Display:

Graphic type LCD, with LED backlight. Text and trace displays.

Operating Temperature:

0 to 50

° C

Ventilation:

Fan cooled

Fan cooled

Operation up to 15,000 feet.

-40 to 75

° C

95% ±5% maximum (non-condensing).

Withstands ±20G, 42ms impulse in X, Y,

Conforms to MIL-STD-167-1.

90 to 260 VAC, 47 to 63 Hz, <50 VA

Altitude:

Operation up to 15,000 feet.

Storage Temperature:

-40 to 75

° C

Humidity:

95% ±5% maximum (non-condensing).

Shock:

Withstands ±20G, 42ms impulse in X, Y, and Z axes, as per MIL-STD-810.

Vibration:

Conforms to MIL-STD-167-1.

Power Requirements:

90 to 260 VAC, 47 to 63 Hz, <50 VA , <30 Watts. No voltage switching
required.

Battery:

One Lithium coin cell for maintaining non-volatile memory information.
Not user replaceable. Typical battery life: 10 years.

Advertising