7 interface, 8 environmental specifications – Boonton 4530 Peak Power Meter User Manual User Manual
Page 19
![background image](/manuals/401796/19/background.png)
Boonton Electronics
Chapter 1
4530 Series RF Power Meter
General Information
1-5
1.4.7 Interface
Video Output:
Compressed representation of detected RF envelope of peak channel(s)
envelope for external oscilloscope monitor or external device synchroni-
zation. This output is roughly logarithmic with input power, is not
calibrated, and can not be used for making any measurements.
Recorder Output:
Programmable voltage output which may be used for monitoring mea-
surements or status of either channel, or for outputting a fixed,
programmable voltage. When used as a measurement monitor, the out-
put is proportial to displayed signal level with programmable or automatic
scaling.
Output range:
0 to 10V (unipolar), or -10V to +10V (bipolar)
Output resolution:
5.0mV
Output impedance:
9K
Absolute accuracy:
±100mV typical, ±200mV max, uncalibrated
±20mV after user calibration
Linearity:
0.1% typical
GPIB Interface:
Complies with IEEE-488.1. Implements AH1, SH1,T6, LE0, SR1, RL1, PP0,
DC1, DT1, C0, and E1.
RS-232 Interface:
Accepts GPIB commands (except bus dependent commands). Provides
for user software updates.
Remote Programming:
SCPI (1990) compliant and Native Mode commands via GPIB or RS-232
interfaces.
Software Drivers:
LABVIEW drivers available.
1.4.8 Environmental Specifications
General:
Manufactured to the intent of MIL-T 28800E, Type III, Class 5, Style E
CE Mark:
Conforms to EU specifications:
EN 61010-1(90)(+A1/92)(+A2/95)
EN 61010-2-031
EN 61326-1(97)
EN 55022(94)(A2/97)Class B
Display:
Graphic type LCD, with LED backlight. Text and trace displays.
Operating Temperature:
0 to 50
° C
Ventilation:
Fan cooled
Fan cooled
Operation up to 15,000 feet.
-40 to 75
° C
95% ±5% maximum (non-condensing).
Withstands ±20G, 42ms impulse in X, Y,
Conforms to MIL-STD-167-1.
90 to 260 VAC, 47 to 63 Hz, <50 VA
Altitude:
Operation up to 15,000 feet.
Storage Temperature:
-40 to 75
° C
Humidity:
95% ±5% maximum (non-condensing).
Shock:
Withstands ±20G, 42ms impulse in X, Y, and Z axes, as per MIL-STD-810.
Vibration:
Conforms to MIL-STD-167-1.
Power Requirements:
90 to 260 VAC, 47 to 63 Hz, <50 VA , <30 Watts. No voltage switching
required.
Battery:
One Lithium coin cell for maintaining non-volatile memory information.
Not user replaceable. Typical battery life: 10 years.