Cw power sensors – Boonton 4530 Peak Power Meter User Manual User Manual

Page 167

Advertising
background image

Boonton Electronics

Appendix A

4530 Series RF Power Meter

Sensors

A-3

CW Power Sensors

Model

Frequency

Dynamic Range

Overload

Maximum SWR

Range

Rating

Impdance

Pulse /

Frequency

SWR

RF Conn

Continuous

Wide Dynamic Range Dual Diode Sensors

51075

500 kHz to 18 GHz

-70 to +20 dBm

1 W for 1

µs

500 kHz to 2 GHz

1.15

50

300 mW

2 GHz to 6 GHz

1.20

N(M)

6 GHz to 8 GHz

1.40

51077

500 kHz to 18 GHz

-60 to +30 dBm

10 W for 1

µs

500 kHz to 2 GHz

1.15

50

3 W

2 GHz to 6 GHz

1.20

N(M)

6 GHz to 18 GHz

1.40

51079

500 kHz to 18 GHz

-50 to +40 dBm

100 W for 1

µs

500 kHz to 2 GHz

1.15

50

25 W

2 GHz to 6 GHz

1.20

N(M)

6 GHz to 18 GHz

1.40

51071

10 MHz to 26.5 GHz

-70 to +20 dBm

1 W for 1

µs

10 MHz to 2 GHz

1.15

50

300 mW

2 GHz to 4 GHz

1.20

K(M)

4 GHz to 18 GHz

1.45

18 GHz to 26.5 GHz

1.50

51072

30 MHz to 40 GHz

-70 to +20 dBm

1 W for 1

µs

30 MHz to 4 GHz

1.25

50

300 mW

4 GHz to 38 GHz

1.65

K(M)

38 GHz to 40 GHz

2.00

Thermocouple Sensors

51100 (9E)

10 MHz to 18 GHz

-20 to +20 dBm

15 W for 1

µs

10 MHz to 30 MHz

1.25

50

300 mW

30 MHz to 16 GHz

1.18

N(M)

16 GHz to 18 GHz

1.28

51101

100 kHz to 4.2 GHz

-20 to +20 dBm

15 W for 1

µs

100 kHz to 300 kHz

1.70

50

300 mW

300 kHz to 2 GHz

1.35

N(M)

2 GHz to 4.2 GHz

1.60

51102

30 MHz to 26.5 GHz

-20 to +20 dBm

150 W for 1

µs

30 MHz to 2 GHz

1.35

50

300 mW

2 GHz to 18 GHz

1.40

K(M)

18 GHz to 26.5 GHz

1.60

51200

10 MHz to 18 GHz

0 to +37 dBm

150 W for 1

µs

10 MHz to 2 GHz

1.10

50

10 W

2 GHz to 12.4 GHz

1.18

N(M)

12.4 GHz to 18 GHz

1.28

51201

100 kHz to 4.2 GHz

0 to +37 dBm

150 W for 1

µs

100 kHz to 2 GHz

1.10

50

10 W

2 GHz to 4.2 GHz

1.18

N(M)

Advertising