21 remote interface command summary, Table 4-1 remote command summary – Boonton 4530 Peak Power Meter User Manual User Manual

Page 130

Advertising
background image

4-56

Chapter 4

Boonton Electronics

Remote Operation

4530 Series RF Power Meter

*TRG

Description:

Trigger a measurement. This command is equivalent to receiving a GET command from the
GPIB bus controller. If the trigger is armed (usually by an

INITiate

command) and trigger

source is set to BUS, a new measurement will begin when

*TRG

is received.

Syntax:

*TRG

Argument:

None

Valid Modes:

Any

4.5.21 Remote Interface Command Summary

The following table is only a summary, and does not contain all command information necessary to pro-
gram the remote interface. See previous the sections in this chapter for detailed command descriptions
and usage syntax. SCPI commands are listed in alphabetical order, followed by native mode commands.

Table 4-1 Remote Command Summary

*CLS

Clear status command to instrument

*ESE

set/return Event Status Enable mask

*ESR?

return value of Event Status Register

*IDN?

return instrument identification string

*OPC[?]

Set OPC flag false, or Return flag when it becomes true

*RST

Global Reset command.

*SRE

set/return value of Service Request Enable mask

*STB?

return status byte, except bit 6 (srq summary) always = 0

*TRG

Trigger a measurement; same action as GET

*TST?

Self test query. Return zero for no errors; signed 16-bit otherwise

*WAI

No-Op for sequential non-overlapped execution

ABORt

immediately set the measurement trigger system to idle

CALCulate[1 | 2]:DCYC

set/return pulse duty cycle, <n> = 0.01 to 100%

CALCulate[1 | 2]:MATH <asc>

set/return the calc math function, <asc> = [see command description]

CALCulate[1 | 2]:MODe <asc>

set/return measurement mode, <asc> = [see command description]

CALCulate[1 | 2]:LIMit:FAIL?

return alarm limit status flags

CALCulate[1 | 2]:LIMit:STATe <asc>

set/return upper and lower limit alarms, <asc> = ON, OFF

CALCulate[1 | 2]:LIMit:CLEar[:IMMediate]

action: reset channel alarm limit status flags

CALCulate[1 | 2]:LIMit:LOWer <n>

set/return the lower limit power level, <n> = -100.00 to +100.00 dBm

CALCulate[1 | 2]:LIMit:UPPer <n>

set/return the upper limit power level, <n> = -100.00 to +100.00 dBm

CALCulate[1 | 2]:PKHLD <asc>

set/return the state of the peak hold function, <asc> = OFF, AVG, INST

CALCulate[1 | 2]:RANGe <asc>

set/return meas range for CW & Voltage sensors, <asc>=AUTO, 0, 1, 2

CALCulate[1 | 2]:REFerence:STATe <asc>

set/return reference measurement mode, <asc> = ON, OFF

CALCulate[1 | 2]:REFerence:COLLect

action: load the sensor’s current power reading as the reference level

CALCulate[1 | 2]:REFerence:DATA <n>

set/return power value of the reference level, <n> = nn.nn dBm

CALCulate[1 | 2]:STATe <asc>

set/return calc channel status, <asc> = ON, OFF

CALCulate[1 | 2]:UNITs <asc>

set/return measurement units, <asc> = [see command description]