4530 Series RF Power Meter

READ:MARKer:POWer

Description:

Performs a single measurement and returns the power reading at the specified marker on the
specified channel.

Syntax:

READ[1|2]:MARKer[1|2]:POWer?

Returns:

power or voltage in active units

Valid Modes:

Pulse and Statistical modes

Restrictions:

MARKer:MODe

must be set to

VERT

READ:ARRay:MARKer:POWer

Description:

Performs a single measurement and returns the power readings at both markers on the
specified channel.

Syntax:

READ[1

|

2]:ARRay:MARKer:POWer?

Returns:

Pwr@Marker1, Pwr@Marker2 in active units

Valid Modes:

Pulse and Statistical modes

Restrictions:

MARKer:MODe

must be set to

VERT

READ:ARRay:MARKer:PERcent

Description:

Performs a single measurement and returns the statistical percent readings at both markers
on the specified channel.

Syntax:

READ[1|2]:ARRay:MARKer:PERcent?

Returns:

Percent@Marker1, Percent@Marker2

Valid Modes:

Statistical mode only

Restrictions:

MARKer:MODe

must be set to

HORZ

READ:ARRay:CW:POWer

Description:

Performs a single measurement and returns the average, maximum, minimum powers and
peak-to-average ratio in dB (peak sensor) or pulse power (CW sensor) of the specified
channel. Note that the values for maximum and minimum power will depend on the peak
hold mode; see the description of the

CALCulate:PKHLD

command for details. If a CW

sensor is used, the pulse power returned is computed from the measured average power
and the preset duty cycle (see

CALCulate:DCYC

command).

Syntax:

READ[1|2]:ARRay:CW:POWer?

Returns:

Pavg, Pmax, Pmin, PkToAvgRatio (Modulated mode) or Ppulse (CW mode)

Valid Modes:

CW and Modulated modes