Boonton 4530 Peak Power Meter User Manual User Manual

Page 89

Advertising
background image

4-15

Boonton Electronics

Chapter 4

4530 Series RF Power Meter

Remote Operation

TKSCWRG

Description:

Returns the CW sensor linearity calibration table from sensor’s EEPROM. CW power
sensors have 7 pairs of upscale / downscale gain factors, and voltage probes have eight
pairs of upscale / midscale / downscale gain factors. Upscale factors have a nominal value
of 5000, and midscale and downscale factors have a nominal value of 0. On voltage probes,
the eighth pair of up/downscale factors and all eight midscale factors is returned at the end
of the array for compatibility with CW sensors.

Syntax:

TKSCWRG?

Returns:

numeric array: < DS0,US0,DS1,US1, ...DS6,US6 [,DS7,US7,MS0,MS1, ...MS7] >

Valid Modes:

All

Restrictions:

SYSTem:LANGuage

must be set to BOON to use native mode instructions

TKSMSG

Description:

Returns the sensor text message from sensor’s EEPROM. This message is programmed at
the factory or during sensor calibration.

Syntax:

TKSMSG?

Returns:

<Message String>

Valid Modes:

All

Restrictions:

SYSTem:LANGuage

must be set to BOON to use native mode instructions

TKAVG

Description:

Sets the talkmode to return the average power of the selected channel each time the 4530 is
addressed to talk. This is a “permanent” talkmode, and will remain in effect until a new
talkmode is set. Use this command in 4530 native mode for the fastest possible sustained
reading rate of single-channel average power over the bus. Power is returned in current
units.

Syntax:

TKAVG

Returns:

Average (or CW) Power

Valid Modes:

CW and Modulated modes

Restrictions:

SYSTem:LANGuage

must be set to BOON to use native mode instructions

Special Case:

If in ratiometric mode, reading will be in units of dBr (log) or Percent Power (linear).

Advertising