Boonton 4530 Peak Power Meter User Manual User Manual
Page 87
4-13
Boonton Electronics
Chapter 4
4530 Series RF Power Meter
Remote Operation
CH1
Description:
Configures native mode talk instructions to return Channel 1 measurement data
Syntax:
CH1
Argument:
None
Valid Modes:
All
Restrictions:
SYSTem:LANGuage
must be set to BOON to use native mode instructions
CH2
Description:
Configures native mode talk instructions to return Channel 2 measurement data
Syntax:
CH2
Argument:
None
Valid Modes:
All
Restrictions:
SYSTem:LANGuag
e must be set to BOON to use native mode instructions
TALKMODE
Description:
Returns the current native mode active channel and talk mode.
Syntax:
TALKMODE?
Returns:
Active channel <CH1, CH2> and talkmode mnemonic
Valid Modes:
All
Restrictions:
SYSTem:LANGuage
must be set to BOON to use native mode instructions
TKERR
Description:
Returns the next queued error code number, 0 if no error. See section 4.9 for a more detailed
description of the error codes that may be returned.
Syntax:
TKERR?
Returns:
<numeric error code>
Valid Modes:
All
Restrictions:
SYSTem:LANGuag
e must be set to BOON to use native mode instructions
TKERRMSG
Description:
Returns the next queued error code number followed by a quoted ASCII text string describ-
ing the error. See section 4.9 for a more detailed description of the error codes that may be
returned.
Syntax:
TKERRMSG?
Returns:
<numeric error code>, “QUOTED ERROR DESCRIPTION”
Valid Modes:
All
Restrictions:
SYSTem:LANGuage must be set to BOON to use native mode instructions