3-36

Chapter 3

Boonton Electronics

Operation

4530 Series RF Power Meter

Main Menu > Calibratr > [ Int Signal, Ext Signal ]

Control the on/off state of the selected calibrator, and modulation source for the External Calibrator.

Off

Disable the selected calibrator’s RF output. (Power-on default)

On CW

Enable selected calibrator’s RF output. Output signal will be unmodulated (CW).

Int/Pulse

Enable external calibrator’s RF output, modulated by its internal pulse generator.

Ext/Pulse

Enable external calibrator’s RF output, modulated by its “ext pulse” input.

Remote Commands:

OUTPut:{INTernal | EXTernal}:SIGNAL <ON, OFF>
OUTPut:EXTernal:MODulation <CW, PULSE>

(controls modulation)

OUTPut:EXTernal:PULSe:SOURce <INT, EXT>

(modulation source)

Main Menu > Calibratr > [ Int Level, Ext Level ]

Set the output level of the selected calibrator in 0.1dB steps.

Range:

-60.0 to +20.0 dBm

Default:

0.0 dBm

Remote Command:

OUTPut:{INTernal | EXTernal}:LEVEL <n>

Main Menu > Calibratr > PlsPeriod

Select the pulse period for the internal pulse modulator of the external calibrator.

Range:

10, 1.0 or 0.1 millisecond (100 Hz, 1kHz or 10kHz)

Default:

10 ms

Remote Command:

OUTPut:EXTernal:PULSe:PERiod <10, 1, 0.1>

Main Menu > Calibratr > DutyCycle

Select the duty cycle for the internal pulse modulator of the external calibrator.

Range:

10% to 90% in 10% steps

Default:

10%

Remote Command:

OUTPut:EXTernal:PULse:DCYC <10, 20, 30, 40, 50, 60, 70, 80, 90)

Main Menu > Calibratr > [ IntStatus, ExtStatus ]

Display a status screen for the selected calibrator. For the internal calibrator, this monitors the oscillator drive
level to verify proper operation. For the external calibrator the software version, serial number, calibration
date, internal temperature and calibration factor are shown.

Remote Command:

None.

Main Menu > Calibratr > [ SelectInt, SelectExt ]

Selects the active calibrator. Only the menu item for the currently inactive calibrator is displayed. Note that
attempting to select the external calibrator will generate an error if it is not connected or turned on.

Remote Command:

No explicit command.

INT

or

EXT

is embedded in any command that requires a

calibrator to be specified.